The FTC300 is a thermal conductivity analyzer with outstanding properties for measuring tasks in process analytics. The device is characterized by high measurement accuracy and stability. It has a comfortable operating menu that allows you to adjust, parameterize and configure without much effort. In addition, operator and data access via an RS232 interface are possible, as well as various analog signal inputs and outputs. The robust design of the FTC300 is designed in protection class IP65 and enables operation at ambient temperatures of -20 ° C to + 50 ° C and is pressure-resistant up to 20 bar. For the measurement of hydrogen in nitrogen, the smallest measuring range is 0 to 0.5 vol.%.
The thermal conductivity analyzer FTC300 can be set up in such a way that switching between different measuring tasks is possible. Concentration determination in up to 16 different binary gas mixtures is possible. This option can be ordered under the keyword “Multi Gas Mode”.
The FTC300 provides a highly effective routine for cross-sensitivity compensation, which can be flexibly adapted to the respective measurement task. It requires as an input external voltage signal, which represents the concentration of the cross-sensitivity causing gas components.
General: | |
Measurement process | Thermal conductivity |
Sampling | Analyzer extractive |
Connection | 6mm pipe |
Dimensions (WxHxD in mm) | 145x80x85 |
Protection class | IP 65 |
Pressure-proof to | 20 bar absolute |
Power supply | 18V to 36V DC / 700mA |
Ambient temperature range | -20°C to 50°C |
Weight | to 1800g |
T90 time at 60l/h | < 1sec |
Communication: | |
RS232 | Yes |
Current output | 1x, 0/4-20 mA |
Voltage output | 2x, 0-10V |
Display | Yes |
Relays | 3x |
Service program SetApp 2.0 | Yes |
Calibration on site | Yes |
Options | |
Flow measurement | Yes |
Flow monitor | Yes |
Infrared measurement | — |
Multi Gas Mode | Yes |
Protection against corrosion | Yes |
Protection against condensate and dust | Yes |
Suitable for inflammable gases | Yes |
Moisture measurement | Yes |
Cross-sensitivity compensation | Yes, external signal required |
Specifications gas analysis | |
Noise | < 1% of smallest range |
Drift at zero point per week | < 2% of smallest range |
Repeatability | < 1% of smallest range |
Non-linearity | < 1% of range |
Measuring error with ambient temperature change per 10°K | < 1% of smallest range |
Flow influence between 60l/h and 90l/h per 10l/h | < 1% of smallest range |
Fault with measurement gas change (Pabs > 800 hPa) per 10 hPa | < 1% of smallest range |
Measuring Gas | Carrier Gas | Basis Range | Smallest Range | Smallest Suppressed Zero Range | Multi Gas Mode MGM |
---|---|---|---|---|---|
Hydrogen (H2) | Nitrogen (N2) or Air | 0% – 100% | 0% – 0.5% | 98% – 100% | Yes |
Oxygen (O2) | Nitrogen (N2) | 0% – 100% | 0% – 15% | 85% – 100% | Yes |
Helium (He) | Nitrogen (N2) or Air | 0% – 100% | 0% – 0.8% | 97% – 100% | Yes |
Carbon dioxide (CO2) | Nitrogen (N2) or Air | 0% – 100% | 0% – 3% | 96% – 100% | Yes |
Nitrogen (N2) | Argon (Ar) | 0% – 100% | 0% – 3% | 97% – 100% | Yes |
Oxygen (O2) | Argon (Ar) | 0% – 100% | 0% – 2% | 97% – 100% | Yes |
Hydrogen (H2) | Argon (Ar) | 0% – 100% | 0% – 0.4% | 99% – 100% | Yes |
Helium (He) | Argon (Ar) | 0% – 100% | 0% – 0.5% | 98% – 100% | Yes |
Carbon dioxide (CO2) | Argon (Ar) | 0% – 60% | 0% – 10% | — | Yes |
Argon (Ar) | Carbon dioxide (CO2) | 40% – 100% | — | 80% – 100% | Yes |
Methane (CH4) | Nitrogen (N2) or Air | 0% – 100% | 0% – 2% | 96% – 100% | Yes |
Methane (CH4) | Argon (Ar) | 0% – 100% | 0% – 1.5% | 97% – 100% | Yes |
Argon (Ar) | Oxygen (O2) | 0% – 100% | 0% – 3% | 96% – 100% | Yes |
Nitrogen (N2) | Hydrogen (H2) | 0% – 100% | 0% – 2% | 99.5% – 100% | Yes |
Oxygen (O2) | Carbon dioxide (CO2) | 0% – 100% | 0% – 3% | 96% – 100% | Yes |
Hydrogen (H2) | Helium (He) | 20% – 100% | 20% – 40% | 85% – 100% | |
Hydrogen (H2) | Methane (CH4) | 0% – 100% | 0% – 0.5% | 98% – 100% | |
Hydrogen (H2) | Carbon dioxide (CO2) | 0% – 100% | 0% – 0.5% | 98% – 100% | |
Sulfur hexafluoride (SF6) | Nitrogen (N2) or Air | 0% – 100% | 0% – 2% | 96% – 100% | |
Nitrogen dioxide (NO2) | Nitrogen (N2) or Air | 0% – 100% | 0% – 5% | 96% – 100% | |
Hydrogen (H2) | Oxygen (O2) | 0% – 100% | 0% – 0.8% | 97% – 100% | |
Argon (Ar) | Xenon (Xe) | 0% – 100% | 0% – 3% | 99% – 100% | |
Neon (Ne) | Argon (Ar) | 0% – 100% | 0% – 1.5% | 99% – 100% | |
Krypton (Kr) | Argon (Ar) | 0% – 100% | 0% – 2% | 96% – 100% | |
Extinguishing gas (R125) | Nitrogen (N2) or Air | 0% – 100% | 0% – 5% | 98% – 100% | |
Deuterium (D2) | Nitrogen (N2) or Air | 0% – 100% | 0% – 0.7% | 96% – 100% | |
Deuterium (D2) | Helium (He) | 0% – 100% | 0% – 5% | 70% – 100% |
phone : +49 (0) 69 530 564 44
fax : +49 (0) 69 530 564 45
e-mail: info@messkonzept.de
When submitting the data entered in the contact form by clicking on the button below, you declare that you have read the privacy policy and that you agree that the data provided will be collected and stored electronically.
The data is used only strictly earmarked for processing and answering your request.
By submitting the contact form you agree to the processing. You may revoke your consent at any time with future effect. In case of cancellation your data will be deleted immediately.