Dimensions without connectors; weight; | 145mm x 80mm x 85mm; max. 1800g |
Power supply | 24V DC (18V to 36V), 700mA |
Ambient temperature range | -5°C to 50°C, other on request |
Linearity | < 1% of range |
Warm up time | About 30min; 1h for small ranges |
Flow rate | 40 l/h to 150 l/h; 60 l/h -80 l/h recommended |
T90-time | < 1 sec at flow rate higher 60 l/h |
Noise | < 1% of smallest range |
Drift at zero point | < 2% of smallest range per week |
Repeatability | < 1% of range |
Error due to change of ambient temperature | < 1% of smallest range per 10°C |
Error due to change of flow at 80 1/h | < 1% of smallest range per 10 l/h |
Gas pressure | Max.2MPa(20 bar) |
Error due to change of pressure (above 800hPa abs.) | < 1% of smallest range oer 10hPa |
Measuring Gas | Carrier Gas | Basis Range | Smallest Range | Smallest Suppressed Zero Range | Multi Gas Mode MGM |
---|---|---|---|---|---|
Measuring GasH2 | Carrier GasN2 or Air | Basis Range0% – 100% | Smallest Range0% – 0.5% | Smallest suppressed zero range98% – 100% | Multi Gas Mode MGMYes |
Measuring GasO2 | Carrier GasN2 | Basis Range0% – 100% | Smallest Range0% – 15% | Smallest suppressed zero range85% – 100% | Multi Gas Mode MGMYes |
Measuring GasHe | Carrier GasN2 or Air | Basis Range0% – 100% | Smallest Range0% – 0.8% | Smallest suppressed zero range97% – 100% | Multi Gas Mode MGMYes |
Measuring GasCO2 | Carrier GasN2 or Air | Basis Range0% – 100% | Smallest Range0% – 3% | Smallest suppressed zero range96% – 100% | Multi Gas Mode MGMYes |
Measuring GasN2 | Carrier GasAr | Basis Range0% – 100% | Smallest Range0% – 3% | Smallest suppressed zero range97% – 100% | Multi Gas Mode MGMYes |
Measuring GasO2 | Carrier GasAr | Basis Range0% – 100% | Smallest Range0% – 2% | Smallest suppressed zero range97% – 100% | Multi Gas Mode MGMYes |
Measuring GasH2 | Carrier GasAr | Basis Range0% – 100% | Smallest Range0% – 0.4% | Smallest suppressed zero range99% – 100% | Multi Gas Mode MGMYes |
Measuring GasHe | Carrier GasAr | Basis Range0% – 100% | Smallest Range0% – 0.5% | Smallest suppressed zero range98% – 100% | Multi Gas Mode MGMYes |
Measuring GasCO2 | Carrier GasAr | Basis Range0% – 60% | Smallest Range0% – 10% | Smallest suppressed zero range— | Multi Gas Mode MGMYes |
Measuring GasAr | Carrier GasCO2 | Basis Range40% – 100% | Smallest Range— | Smallest suppressed zero range80% – 100% | Multi Gas Mode MGMYes |
Measuring GasCH4 | Carrier GasN2 or Air | Basis Range0% – 100% | Smallest Range0% – 2% | Smallest suppressed zero range96% – 100% | Multi Gas Mode MGMYes |
Measuring GasCH4 | Carrier GasAr | Basis Range0% – 100% | Smallest Range0% – 1.5% | Smallest suppressed zero range97% – 100% | Multi Gas Mode MGMYes |
Measuring GasAr | Carrier GasO2 | Basis Range0% – 100% | Smallest Range0% – 3% | Smallest suppressed zero range96% – 100% | Multi Gas Mode MGMYes |
Measuring GasN2 | Carrier GasH2 | Basis Range0% – 100% | Smallest Range0% – 2% | Smallest suppressed zero range99.5% – 100% | Multi Gas Mode MGMYes |
Measuring GasO2 | Carrier GasCO2 | Basis Range0% – 100% | Smallest Range0% – 3% | Smallest suppressed zero range96% – 100% | Multi Gas Mode MGMYes |
Measuring GasH2 | Carrier GasHe | Basis Range20% – 100% | Smallest Range20% – 40% | Smallest suppressed zero range85% – 100% | Multi Gas Mode MGM |
Measuring GasH2 | Carrier GasCH4 | Basis Range0% – 100% | Smallest Range0% – 0.5% | Smallest suppressed zero range98% – 100% | Multi Gas Mode MGM |
Measuring GasH2 | Carrier GasCO2 | Basis Range0% – 100% | Smallest Range0% – 0.5% | Smallest suppressed zero range98% – 100% | Multi Gas Mode MGM |
Measuring GasSF6 | Carrier GasN2 or Air | Basis Range0% – 100% | Smallest Range0% – 2% | Smallest suppressed zero range96% – 100% | Multi Gas Mode MGM |
Measuring GasNO2 | Carrier GasN2 or Air | Basis Range0% – 100% | Smallest Range0% – 5% | Smallest suppressed zero range96% – 100% | Multi Gas Mode MGM |
Measuring GasH2 | Carrier GasO2 | Basis Range0% – 100% | Smallest Range0% – 4% | Smallest suppressed zero range97% – 100% | Multi Gas Mode MGM |
Measuring GasAr | Carrier GasXe | Basis Range0% – 100% | Smallest Range0% – 3% | Smallest suppressed zero range99% – 100% | Multi Gas Mode MGM |
Measuring GasNe | Carrier GasAr | Basis Range0% – 100% | Smallest Range0% – 1.5% | Smallest suppressed zero range99% – 100% | Multi Gas Mode MGM |
Measuring GasKr | Carrier GasAr | Basis Range0% – 100% | Smallest Range0% – 2% | Smallest suppressed zero range96% – 100% | Multi Gas Mode MGM |
Measuring GasR125 | Carrier GasN2 or Air | Basis Range0% – 100% | Smallest Range0% – 4% | Smallest suppressed zero range98% – 100% | Multi Gas Mode MGM |
Contact Center : ( +49 ) 69 530 564 44
fax : ( +49 ) 69 530 564 45
e-mail: info@messkonzept.de
Contact Center :
( +49 ) 69 530 564 44