The FTC200-OEM is a highly precise and stable Thermal Conductivity Detector (TCD). It is designed for the use as an OEM detector. Since it is not equipped with housing it has protection class 00 and no electro-magnetic shielding. It is not designed for the use in hazardous areas.
For indication of the signal a 0 to 10V analog output is provided. For internal calibration, configuration and digital indication of the signal a TTL-level RS232 communication device is required. For testing purposes we offer a RS232 to TTL level converter as well as two small windows programs for any PC in order to perform offset and gain calibration.
In given cases, a cross sensitive gas may be measured separately and the signal of this measurement is fed into FTC200 where an internal routine compensates the cross sensitivity.
General: | |
Measurement process | Thermal conductivity |
Sampling | OEM |
Connection | 6mm pipe |
Dimensions (WxHxD in mm) | 90x40x40 |
Protection class | None |
Pressure-proof to | 20 bar absolut |
Power supply | 18V to 36V DC / 700mA |
Ambient temperature range | -20°C to 50°C |
Weight | to 200 g |
T90 time at 60l/h | < 1sec |
Communication: | |
RS232 | Yes |
Current output | — |
Voltage output | 2x, 0-10V |
Display | — |
Relays | — |
Service program SetApp 2.0 | Yes |
Calibration on site | Yes |
Options | |
Flow measurement | Yes |
Flow monitor | Yes |
Infrared measurement | — |
Multi Gas Mode | Yes |
Protection against corrosion | Yes |
Protection against condensate and dust | Yes |
Suitable for inflammable gases | customer responsibility |
Moisture measurement | Yes |
Cross-sensitivity compensation | Yes, external signal required |
Specifications gas analysis | |
Noise | < 1% of smallest range |
Drift at zero point per week | < 2% of smallest range |
Repeatability | < 1% of smallest range |
Non-linearity | < 1% of range |
Measuring error with ambient temperature change per 10°K | < 1% of smallest range |
Flow influence between 60l/h and 90l/h per 10l/h | < 1% of smallest range |
Fault with measurement gas change (Pabs > 800 hPa) per 10 hPa | < 1% of smallest range |
Measuring Gas | Carrier Gas | Basis Range | Smallest Range | Smallest Suppressed Zero Range | Multi Gas Mode MGM |
---|---|---|---|---|---|
Hydrogen (H2) | Nitrogen (N2) or Air | 0% – 100% | 0% – 0.5% | 98% – 100% | Yes |
Oxygen (O2) | Nitrogen (N2) | 0% – 100% | 0% – 15% | 85% – 100% | Yes |
Helium (He) | Nitrogen (N2) or Air | 0% – 100% | 0% – 0.8% | 97% – 100% | Yes |
Carbon dioxide (CO2) | Nitrogen (N2) or Air | 0% – 100% | 0% – 3% | 96% – 100% | Yes |
Nitrogen (N2) | Argon (Ar) | 0% – 100% | 0% – 3% | 97% – 100% | Yes |
Oxygen (O2) | Argon (Ar) | 0% – 100% | 0% – 2% | 97% – 100% | Yes |
Hydrogen (H2) | Argon (Ar) | 0% – 100% | 0% – 0.4% | 99% – 100% | Yes |
Helium (He) | Argon (Ar) | 0% – 100% | 0% – 0.5% | 98% – 100% | Yes |
Carbon dioxide (CO2) | Argon (Ar) | 0% – 60% | 0% – 10% | — | Yes |
Argon (Ar) | Carbon dioxide (CO2) | 40% – 100% | — | 80% – 100% | Yes |
Methane (CH4) | Nitrogen (N2) or Air | 0% – 100% | 0% – 2% | 96% – 100% | Yes |
Methane (CH4) | Argon (Ar) | 0% – 100% | 0% – 1.5% | 97% – 100% | Yes |
Argon (Ar) | Oxygen (O2) | 0% – 100% | 0% – 3% | 96% – 100% | Yes |
Nitrogen (N2) | Hydrogen (H2) | 0% – 100% | 0% – 2% | 99.5% – 100% | Yes |
Oxygen (O2) | Carbon dioxide (CO2) | 0% – 100% | 0% – 3% | 96% – 100% | Yes |
Hydrogen (H2) | Helium (He) | 20% – 100% | 20% – 40% | 85% – 100% | |
Hydrogen (H2) | Methane (CH4) | 0% – 100% | 0% – 0.5% | 98% – 100% | |
Hydrogen (H2) | Carbon dioxide (CO2) | 0% – 100% | 0% – 0.5% | 98% – 100% | |
Sulfur hexafluoride (SF6) | Nitrogen (N2) or Air | 0% – 100% | 0% – 2% | 96% – 100% | |
Nitrogen dioxide (NO2) | Nitrogen (N2) or Air | 0% – 100% | 0% – 5% | 96% – 100% | |
Hydrogen (H2) | Oxygen (O2) | 0% – 100% | 0% – 0.8% | 97% – 100% | |
Argon (Ar) | Xenon (Xe) | 0% – 100% | 0% – 3% | 99% – 100% | |
Neon (Ne) | Argon (Ar) | 0% – 100% | 0% – 1.5% | 99% – 100% | |
Krypton (Kr) | Argon (Ar) | 0% – 100% | 0% – 2% | 96% – 100% | |
Extinguishing gas (R125) | Nitrogen (N2) or Air | 0% – 100% | 0% – 5% | 98% – 100% | |
Deuterium (D2) | Nitrogen (N2) or Air | 0% – 100% | 0% – 0.7% | 96% – 100% | |
Deuterium (D2) | Helium (He) | 0% – 100% | 0% – 5% | 70% – 100% |
phone : +49 (0) 69 530 564 44
fax : +49 (0) 69 530 564 45
e-mail : info@messkonzept.de
When submitting the data entered in the contact form by clicking on the button below, you declare that you have read the privacy policy and that you agree that the data provided will be collected and stored electronically.
The data is used only strictly earmarked for processing and answering your request.
By submitting the contact form you agree to the processing. You may revoke your consent at any time with future effect. In case of cancellation your data will be deleted immediately.